Calculate accelerated aging time for a target shelf life, or the real-time aging a chamber test represents, using the Q10 (ASTM F1980) or Arrhenius model.

Estimate chamber time for a target real-time duration.

Typical room-temperature storage is 25 ยฐC (ASTM F1980 uses 20โ€“25 ยฐC). Change it to your storage condition.

55 ยฐC is a common chamber setting; ASTM F1980 advises staying at or below 60 ยฐC for most medical packaging.

2 is the conventional conservative value (ASTM F1980). Use a product-specific value if you have one.

Model month = 30.4375 days; year = 365.25 days.

Optional chamber schedule
Blank means no added margin; a margin does not validate the model.

Blank omits the end date.


Related Calculators

Calculator Functions

FunctionWhat it doesWhat it requires
Required chamber timeFinds the days a sample must spend at the accelerated temperature to stand for a target real-time durationRequires the storage temperature, the accelerated test temperature, a Q10 factor or an activation energy, and the target real-time duration
Equivalent real-time agingFinds the real-time aging that a completed chamber run stands forRequires the storage temperature, the accelerated test temperature, a Q10 factor or an activation energy, and the completed chamber duration

Both functions use the Q10 model or the Arrhenius model, chosen under Acceleration model. The storage temperature starts at 25 ยฐC, the test temperature at 55 ยฐC and Q10 at 2. Durations go in as days, weeks, months of 30.4375 days or years of 365.25 days. Required chamber time has two optional settings: an added chamber-time margin in percent, and a chamber start date that gives an estimated end date. Every result also shows the acceleration factor.

Accelerated Stability Formulas

AAF = Q10(TAA โˆ’ TRT) รท 10 [1]
AAFAccelerated aging factor: how many times faster aging runs in the chamber than in storage
Q10Factor by which the aging rate rises for each 10 ยฐC increase. Values usually fall between 1.8 and 2.5, and 2.0 is the most common choice [1]
TAAAccelerated test temperature (ยฐC)
TRTReal-time storage temperature (ยฐC)

This is the form used for accelerated aging of medical devices and their sterile packaging [1]. Storage temperatures of 20 ยฐC to 25 ยฐC and chamber temperatures of 50 ยฐC, 55 ยฐC or 60 ยฐC are typical for that work [1]. The test temperature must be higher than the storage temperature, and Q10 must be greater than 1.

Required chamber time

AAT = tRT รท AAF [1]

AAT is the accelerated aging time in days and tRT is the target real-time duration in days. For 12 months at 25 ยฐC, tested at 55 ยฐC with Q10 = 2, AAF = 23 = 8 and AAT = 365.25 รท 8 = 45.66 days. An added margin multiplies that figure by (1 + margin รท 100), and the answer is rounded up to whole chamber days, 46 here. With a start date, the end date is the start date plus that whole number of days.

Equivalent real-time aging

tRT = AAT ร— AAF

This is the chamber-time formula turned around. A completed run of 46 days at 55 ยฐC, for storage at 25 ยฐC with Q10 = 2, stands for 46 ร— 8 = 368 days of real-time aging. The answer is given in the unit the chamber duration was entered in.

Arrhenius model

AAF = exp[(Ea รท R) ร—
(1 รท TRT โˆ’ 1 รท TAA)] [2]

With the Arrhenius model selected, the factor comes from an activation energy in place of Q10. Ea is the activation energy in J/mol (the entry in kJ/mol ร— 1,000), R is the molar gas constant, 8.314462618 J/(molยทK) [3], and both temperatures are in kelvin (ยฐC + 273.15). The expression is the ratio of the reaction rate constants at the two temperatures [2]. With Ea = 50 kJ/mol, 25 ยฐC storage and a 55 ยฐC chamber, AAF = 6.3215, so 12 months needs 365.25 รท 6.3215 = 57.78 days, rounded up to 58. Either function then uses this factor in the same way as the Q10 factor.

Limits of This Estimate

  • The result is an estimated equivalence. It does not establish product quality, safety or an approved expiry date. For medical devices, accelerated aging data supports a tentative expiration date while real-time shelf-life studies run to confirm it [1].
  • Both models assume one degradation mechanism whose rate depends only on temperature. Heat can damage some plastics, adhesives and components in ways that never happen in storage, and humidity is not part of the equation [1].
  • Q10 = 2 is a starting assumption, not a measured property of a product. A Q10 or activation energy taken from stability data for the product itself gives a better estimate, and an added margin does not validate the model.

Sources

  1. WESTPAK, Three Critical Items You Must Know to Plan and Start Accelerated Aging for Medical Devices (2023): accelerated aging time formula used with ASTM F1980, typical Q10 and temperatures, and the need for real-time studies
  2. University of Calgary, Chemistry Textbook, Activation Energy and the Arrhenius Equation: Arrhenius equation and its two-temperature form
  3. NIST, CODATA Value: molar gas constant: value of R