Correct brewing hydrometer SG, calculate free-air and Bouguer anomalies, or adjust indicated mass for local gravity and air buoyancy with visible terms.
- All Physics Calculators
- Specific Gravity Calculator
- Specific Gravity to Density Calculator
- Plato to SG Calculator
- Brix Density Calculator
- Liquid Density Calculator
Brewing Temperature Correction Formula
The following formula approximates a hydrometer temperature correction in Specific Gravity (SG), assuming water-like liquid expansion and neglecting hydrometer glass expansion. This calculator accepts and returns SG; it does not convert ยฐPlato.
Variables:
- SGcorr is the temperature-corrected specific gravity (dimensionless)
- SGmeas is the measured specific gravity at the sample temperature (dimensionless)
- ฯw(T) is the density of pure water at temperature T (kg/mยณ)
- Tmeas is the sample temperature (ยฐC)
- Tref is the instrument calibration/reference temperature (ยฐC)
Note: hydrometer temperature correction can be instrument-specific (and may include glass expansion and wort thermal expansion). For best accuracy, use the manufacturerโs correction table. This calculator uses a water-density approximation over a numerical temperature range of 0โ100ยฐC. That range does not establish hydrometer accuracy; prefer measurement near calibration temperature and the manufacturerโs instrument/liquid correction table.
What is Gravity Correction?
โGravity correctionโ can mean different things depending on context. In brewing, it usually means adjusting a hydrometer gravity reading for temperature differences from the hydrometerโs calibration temperature (commonly 20ยฐC or 60ยฐF). In geophysics, gravity corrections are applied to measured gravitational acceleration to help compute anomalies (such as free-air and Bouguer anomalies). In mass metrology, instrument calibration determines whether a local-gravity correction is needed. A locally calibrated comparison cancels gravity, while the modeled force-responsive scale calibrated against reference weights at standard gravity requires the stated g0/local-gravity factor. Both mass paths assume identical air density during calibration and use.
How to Calculate Corrected Gravity?
The following steps outline how to calculate a brewing temperature-corrected gravity reading.
- Measure the hydrometer reading as Specific Gravity (SG).
- Record the sample temperature (Tmeas) and the hydrometerโs calibration/reference temperature (Tref), using consistent units (e.g., ยฐC).
- If your measurement is in ยฐPlato, use a suitable separate conversion before entering SG; no automatic Plato conversion is provided here.
- Compute the density of pure water at Tmeas and Tref, then apply: SGcorr = SGmeas ร ฯw(Tref)/ฯw(Tmeas).
- If desired, convert SGcorr back to ยฐPlato.
- Choose Brewing hydrometer temperature in the Solve for selector, then Calculate.
Example Problem:
Use the following variables as an example problem to test your knowledge.
measured specific gravity = 1.05 SG
sample temperature (Tmeas) = 25 ยฐC
calibration temperature (Tref) = 20 ยฐC
