Correct an observed refractive index to 20 °C or another reference temperature with a typical or custom correction factor, or estimate the factor.
- All Physics Calculators
- Effective Refractive Index Calculator
- Lens Index Calculator
- Spherical Equivalent Calculator
Calculator Functions
| Function | What it does | What it requires |
|---|---|---|
| Corrected refractive index | Finds the refractive index at the reference temperature from a reading taken at another temperature | Requires the observed refractive index, the observed temperature, the reference temperature and a correction factor |
| Observed refractive index | Finds the reading to expect at the measuring temperature from a reference value | Requires the corrected refractive index, the observed temperature, the reference temperature and a correction factor |
| Observed temperature | Finds the temperature at which a reading was taken | Requires the observed and corrected refractive indexes, the reference temperature and a correction factor |
| Reference temperature | Finds the temperature a corrected value refers to | Requires the observed and corrected refractive indexes, the observed temperature and a correction factor |
| Correction factor from observed/reference values | Finds the correction factor that links an observed reading to a reference value | Requires the observed and corrected refractive indexes, the observed temperature and the reference temperature |
| Correction factor from two measurements | Finds the correction factor of a liquid from readings at two temperatures | Requires the refractive index and temperature of each of the two measurements |
Temperatures are entered in °C, °F or K, and the reference temperature starts at 20 °C with a 25 °C shortcut. The correction factor is chosen from 0.00045 per °C for typical organic liquids, 0.0001 per °C for water and dilute aqueous solutions, or a custom value per °C, per °F or per K.
Refractive Index Correction Formula
CF × (Tobs − Tref) [1]
| nref | Refractive index corrected to the reference temperature |
| nobs | Refractive index read at the observed temperature |
| CF | Correction factor, the fall in refractive index for each 1 °C rise in temperature |
| Tobs | Temperature of the sample when it was read (°C) |
| Tref | Reference temperature, usually 20 °C |
The refractive index of a liquid falls as it warms, so a reading taken above the reference temperature is corrected upward and a reading taken below it is corrected downward. For most organic liquids the fall is about 0.00045 for every 1 °C, give or take 0.0001, and for water it is much smaller, about 0.0001 per °C [1]. A factor entered per °F is multiplied by 1.8 to put it on a per °C basis.
Corrected refractive index
The main formula applies directly. An organic liquid that reads 1.4370 at 18 °C corrects to 1.4370 + 0.00045 × (18 − 20) = 1.4361 at 20 °C [3]. A reading of 1.3330 at 25 °C with the same factor corrects to 1.3330 + 0.00045 × 5 = 1.33525. With the water factor, 1.3325 at 25 °C corrects to 1.3325 + 0.0001 × 5 = 1.3330, which agrees with the measured index of pure water, 1.33250 at 25 °C and 1.33299 at 20 °C [2].
Observed refractive index
CF × (Tobs − Tref)
A liquid listed at 1.33525 for 20 °C should read 1.33525 − 0.00045 × (25 − 20) = 1.3330 on a refractometer at 25 °C.
Observed temperature
(nref − nobs) ÷ CF
A reading of 1.3330 for a liquid listed at 1.33525 for 20 °C, with a factor of 0.00045, was taken at 20 + 0.00225 ÷ 0.00045 = 25 °C.
Reference temperature
(nref − nobs) ÷ CF
A reading of 1.3330 at 25 °C that corrects to 1.33525 with a factor of 0.00045 refers to 25 − 0.00225 ÷ 0.00045 = 20 °C.
Correction factor from observed/reference values
(Tobs − Tref)
An observed 1.3330 at 25 °C against a reference value of 1.33525 at 20 °C gives (1.33525 − 1.3330) ÷ 5 = 0.00045 per °C. The two temperatures must differ.
Correction factor from two measurements
n1 and n2 are readings of the same liquid at temperatures T1 and T2. Pure water measures 1.3329877 at 20 °C and 1.3325026 at 25 °C [2], so its factor over that range is (1.3329877 − 1.3325026) ÷ 5 = 0.000097 per °C. A factor measured this way for the actual sample is better than either typical value.
Limits of This Estimate
- The correction is a straight-line approximation. The typical organic factor varies by about ±0.0001 per °C from one liquid to another, and it should not be used for water-based solutions [1].
- The factor itself changes with temperature. For water it is about 0.00009 per °C at 20 °C and 0.00011 per °C at 25 °C [2], so corrections over a wide temperature range lose accuracy.
- Both values must be for the same wavelength, normally the sodium D line at 589.3 nm, and a typical laboratory refractometer reads to about ±0.0002 [1]. Correcting for temperature does not improve on that.
Sources
- University of Puget Sound, Chemistry Lab Techniques, Refractometry: Analyzing Results: temperature correction, the 0.00045 and 0.0001 per °C factors, wavelength and instrument precision
- Tilton and Taylor, Refractive Index and Dispersion of Distilled Water for Visible Radiation, at Temperatures 0 to 60 °C, Journal of Research of the National Bureau of Standards, RP1085 (1938), Table 6: measured refractive index of water at each temperature
- Meredith College, Organic I Techniques, 10: Refractive Index: worked correction of a reading taken at 18 °C to 20 °C
