Correct a refractometer reading to 20 °C using a typical temperature coefficient for water or essential oils, or enter your sample’s own coefficient.

Standard reference is 20 °C. Use your method’s reference if it differs.

CF = −dnD/dT for your sample; its sign matters.

Optional settings

Calculator Functions

FunctionWhat it doesWhat it requires
Correct observed refractive indexConverts a refractive index read at the sample temperature to its value at the reference temperatureRequires the observed refractive index, the sample temperature, the reference temperature and a sample type or custom coefficient
Recover observed refractive indexFinds the reading expected at the sample temperature from a reference-temperature valueRequires the reference-temperature refractive index, the sample temperature, the reference temperature and a sample type or custom coefficient
Find correction coefficientFinds the correction coefficient from readings at two temperaturesRequires the observed refractive index, the reference-temperature refractive index and both temperatures
Find sample temperatureFinds the sample temperature that explains the gap between two readingsRequires the observed and reference-temperature refractive indexes, the reference temperature, a sample type or custom coefficient and the result unit

Temperatures are entered in °C, °F or K, and the reference temperature starts at 20 °C. The sample type sets the coefficient: about 0.0001 nD/°C for water and dilute water solutions, 0.0004 nD/°C for essential oils, or a custom value. Every result lists both refractive indexes, both temperatures in °C and the correction term.

Refractometer Temperature Correction Formulas

nref = nobs + CF ×
(Ts − Tref) [1]
nrefRefractive index nD at the reference temperature
nobsRefractive index nD read at the sample temperature
CFCorrection coefficient (nD per °C), the drop in refractive index for each 1 °C rise
TsSample temperature (°C)
TrefReference temperature (°C), usually 20 °C

The refractive index of a liquid falls as it warms, so a reading taken warm is lower than the 20 °C value and the correction adds to it. One refractometer manual defines the slope to enter as the index at the lower temperature minus the index at the higher temperature, divided by the temperature difference, and requires it to be known for the specific sample [1]. Its example is a sample reading 1.3334 at 15 °C and 1.3319 at 30 °C, a slope of 0.0001 per °C [1].

Correct observed refractive index

The main formula applies directly. A water sample reading 1.3319 at 30 °C, corrected to 20 °C with the 0.0001 coefficient, is 1.3319 + 0.0001 × (30 − 20) = 1.3329.

Recover observed refractive index

nobs = nref − CF ×
(Ts − Tref)

A liquid with a 20 °C value of 1.3329 and a coefficient of 0.0001 should read 1.3329 − 0.0001 × 10 = 1.3319 at 30 °C.

Find correction coefficient

CF = (nref − nobs) ÷
(Ts − Tref)

With 1.3334 at a 15 °C reference and 1.3319 observed at 30 °C, CF = (1.3334 − 1.3319) ÷ (30 − 15) = 0.0001 nD/°C [1]. The two temperatures must differ.

Find sample temperature

Ts = Tref +
(nref − nobs) ÷ CF

A reading of 1.3319 against a 20 °C value of 1.3329, at 0.0001 nD/°C, points to a sample at 20 + 0.0010 ÷ 0.0001 = 30 °C. The coefficient cannot be zero.

Limits of This Estimate

  • A single coefficient is a straight-line approximation over a small temperature range. The coefficient has to match the sample, its concentration and the wavelength of the instrument [1].
  • The two sample-type presets are typical values, not measured ones for a given sample. These estimates are based on the best available data and estimations.
  • This is not a Brix correction. Refractometers correct Brix readings with a separate sucrose temperature table [1], and no sucrose coefficient is built in here.
  • A reading that the instrument has already temperature-compensated should not be corrected again.

Sources

  1. METTLER TOLEDO, RE40 Refractometer Operating Instructions, Section 9.2: the slope definition, the worked 0.0001 example, the sample-specific requirement and the separate Brix compensation