Calculate absorbance, transmittance, and % light passing through a sample from Beer-Lambert inputs or intensity readings, with AU and %T outputs.

Absorbance to Transmittance Calculator

Choose the form that matches what you have, then calculate.

Absorbance → %T
%T → Absorbance
Beer-Lambert
Light readings

Absorbance to Transmittance Formula

T = 10^(-A)    %T = 10^(-A) * 100
  • A = absorbance (AU, unitless)
  • T = transmittance as a decimal fraction (0 to 1)
  • %T = transmittance as a percentage (0 to 100)

The reverse direction uses A = -log₁₀(T). If you only have raw light readings, T = I/I₀ where I is the transmitted intensity and I₀ is the reference. To predict A from sample properties, the calculator applies the Beer-Lambert law A = εlc, then converts to %T. Path length is converted to cm and concentration to mol/L before multiplying.

Reference Tables

Quick conversions between absorbance and transmittance:

Absorbance (A)Transmittance (%T)Light blocked
0.00100%0%
0.1079.4%20.6%
0.3050.1%49.9%
0.5031.6%68.4%
1.0010.0%90.0%
1.503.16%96.84%
2.001.00%99.00%
3.000.10%99.90%

How to interpret a reading:

RangeWhat it means
A < 0.1Signal is weak; consider concentrating the sample.
A 0.1 to 1.0Standard linear range for most spectrophotometers.
A 1.0 to 2.0Approaching detector limits; dilute if possible.
A > 2.0Stray light dominates; results unreliable.
Negative ASample reading exceeds reference; rerun the blank.

Worked Example and FAQ

Example. A sample reads A = 0.75 at 540 nm. Then T = 10⁻⁰·⁷⁵ = 0.1778, so %T = 17.78%. About 17.78% of the incident light passes through; the rest is absorbed or scattered.

Why is absorbance unitless? It is the log ratio of two intensities (I₀/I), so the units cancel. The "AU" label just marks that it came from an absorbance measurement.

Can transmittance exceed 100%? Not for a true absorbing sample. A %T above 100 means the reference reading was lower than the sample reading, usually a blanking error or a fluorescent sample.

Which form should I use? Use Absorbance → %T or %T → Absorbance for direct conversion. Use Beer-Lambert when you know ε, path length, and concentration. Use Light readings when you have raw I and I₀ from a detector.

absorbance to transmittance calculator
absorbance to transmittance formula