Calculate normalized intensity from sample, background and reference light readings in lux, foot-candles or cd/m², or solve for any one missing reading.

All three readings must measure the same quantity. Linear normalization does not model perceived brightness.

Optional settings


Related Calculators

Normalized Intensity Formula

The normalized intensity compares a sample intensity to a background and reference intensity. The main formula is:

NI = (Is - Ib)/(Ir - Ib)

The calculator can also rearrange the formula to solve for any missing value:

Is = NI×(Ir - Ib) + Ib
Ib = (Is - NI×Ir)/(1 - NI)
Ir = ((Is - Ib)/NI) + Ib
  • NI = normalized intensity, unitless
  • Is = intensity of sample
  • Ib = intensity of background
  • Ir = intensity of reference

To calculate normalized intensity, enter the sample, background, and reference intensities. To solve for one of the intensity values, enter normalized intensity plus the other two intensity values.

Illuminance readings are converted to lux using 1 fc = 1 / 0.09290304 lx (about 10.76391 lx). Luminance readings remain in cd/m². Lux and cd/m² measure different quantities and cannot generally be converted without geometry or a reflection model.

Intensity Unit Conversions and Normalized Result Ranges

Unit Symbol Conversion used by calculator
Lux lx 1 lx = 1 lx
Foot-candle fc 1 fc = 10.7639 lx
Candela per square meter cd/m² Luminance only; no general conversion to lux
Normalized intensity Meaning (reference above background)
0 The sample equals the background intensity.
Between 0 and 1 The sample is between the background and reference intensities.
1 The sample equals the reference intensity.
Less than 0 The sample is below the background intensity.
Greater than 1 The sample is above the reference intensity.

Examples

Example 1: Calculate normalized intensity

You have a sample intensity of 70 lx, a background intensity of 20 lx, and a reference intensity of 120 lx.

NI = (70 - 20) / (120 - 20)
NI = 50 / 100 = 0.5

The normalized intensity is 0.5000.

Example 2: Calculate sample intensity

You have a normalized intensity of 0.75, a background intensity of 10 lx, and a reference intensity of 90 lx.

Is = 0.75×(90 - 10) + 10
Is = 0.75×80 + 10 = 70

The sample intensity is 70 lx.

FAQ

Can normalized intensity be greater than 1?

Yes. A normalized intensity greater than 1 means the sample intensity is higher than the reference intensity, assuming the reference is greater than the background.

Can normalized intensity be negative?

Yes. A negative value means the sample intensity is lower than the background intensity, assuming the reference is greater than the background.

Why can the reference and background not be the same?

The formula divides by the difference between the reference and background intensities. If they are the same, the denominator is zero, so normalized intensity cannot be calculated.